Ultrasound: eFAST, aorta and vascular access

Point-of-care ultrasound to detect free fluid, pericardial fluid, pneumothorax and haemothorax in trauma (eFAST), screen the abdominal aorta for aneurysm, and guide vascular access.

Open in the appFree, works on your phone and offline, with every tool alongside.

Do not forget

Indications

Contraindications

Absolute

Relative

Equipment

Key numbers

Aneurysm threshold
Abdominal aortic diameter 3 cm or more
Large aneurysm
5.5 cm or more — refer to vascular surgery
Aorta measurement
Outer wall to outer wall, anterior–posterior, transverse view
eFAST views
RUQ, LUQ, pelvis, subxiphoid cardiac, bilateral anterior thorax
Cannula in vein
At least two-thirds of the cannula length

Before · Preparation

  1. Decide the clinical question you are answering (e.g. 'Is there free fluid?', 'Is there an aneurysm?').
  2. Introduce yourself, explain the scan and obtain verbal consent where possible.
  3. Enter patient details on the machine so images can be stored.
  4. Select the correct probe and preset; orientate the probe marker to the patient's right or head, matching the screen marker.
  5. Position the patient supine; dim the lights if possible.
  6. Perform hand hygiene and clean the probe.

During · eFAST — abdomen

  1. Right upper quadrant: curvilinear probe in the right mid- to posterior axillary line at the 8th–11th rib spaces, marker to the head; view the hepatorenal space (Morison's pouch), including the caudal tip of the liver.The liver tip is the most sensitive area for free fluid.
  2. Look above the diaphragm in the same view for haemothorax (anechoic fluid, loss of mirror image, visible spine above the diaphragm).
  3. Left upper quadrant: probe more posterior and higher, posterior axillary line; view the splenorenal space and the subphrenic space between spleen and diaphragm, and above the diaphragm for haemothorax.
  4. Pelvis: probe just above the pubic symphysis in transverse and sagittal planes; look behind the bladder (rectovesical pouch in men, pouch of Douglas in women).A full bladder helps; scan before catheterisation if possible.

During · eFAST — heart and lungs

  1. Subxiphoid cardiac view: probe flat under the xiphisternum aimed at the left shoulder; look for pericardial fluid (anechoic stripe) and cardiac activity; use the parasternal long-axis view if subxiphoid is poor.
  2. Anterior lungs: linear or curvilinear probe in the second or third intercostal space, mid-clavicular line, then lower anterior and lateral points on each side.
  3. Look for lung sliding at the pleural line and B-lines; use M-mode to see the 'seashore' (normal) or 'barcode' pattern (no sliding).
  4. Look for a lung point, which confirms pneumothorax.Absent sliding alone can also be due to intubation of the other main bronchus, adhesions or apnoea.
  5. Repeat the eFAST if the patient's condition changes.

During · Aorta

  1. Curvilinear probe in the epigastrium, transverse, marker to the patient's right; identify the vertebral body and the aorta anterior and slightly to the left of it (inferior vena cava to the right).
  2. Follow the aorta distally from the coeliac axis and superior mesenteric artery to the bifurcation near the umbilicus, then into the iliac arteries.
  3. Measure the maximum diameter in transverse view from outer wall to outer wall (anterior–posterior), and confirm in longitudinal view.
  4. An abdominal aortic diameter of 3 cm or more is an aneurysm.
  5. Do not use ultrasound to exclude rupture: a patient with an aneurysm and hypotension, pain or collapse must be referred immediately to vascular surgery.Retroperitoneal haemorrhage is usually not visible on ultrasound.

During · Vascular access

  1. Use the linear probe; scan the target vessel and identify vein (compressible, non-pulsatile, thin-walled) and artery (pulsatile, non-compressible); use colour Doppler if unsure.
  2. Choose a vein that is straight, of adequate size and not too deep for the cannula length (at least two-thirds of the cannula should lie within the vein).
  3. For procedures, use full asepsis with a sterile probe cover and sterile gel.
  4. Out of plane (short axis): centre the vein on the screen, insert the needle at the probe midpoint and 'walk' the tip down by sliding the probe as the needle advances.
  5. In plane (long axis): align the needle along the length of the probe so the whole shaft and tip are visible.
  6. Keep the needle tip in view at all times; confirm the tip inside the vessel before advancing the cannula or guidewire.

After · Interpretation and action

  1. State findings as positive, negative or indeterminate for each view, and act on the clinical question.
  2. A negative FAST does not exclude intra-abdominal, retroperitoneal, bowel or diaphragmatic injury.Use CT in the stable patient and repeat the scan if the condition changes.
  3. In the unstable trauma patient with a positive FAST, proceed to haemorrhage control (theatre or interventional radiology) without further imaging, per senior decision.

After · Documentation and cleaning

  1. Save representative images or clips of each view.
  2. Document the indication, views obtained, findings, limitations and actions.
  3. Clean and disinfect the probe after each patient; use high-level disinfection after use on non-intact skin or with a sterile sheath breach, per policy.

Complications

Pitfalls

In children

FAST is less sensitive in children because solid organ injuries often occur without free fluid; a negative FAST must not exclude injury, and CT or observation decisions follow paediatric trauma guidance. Use higher-frequency probes and ultrasound-guided vascular access with appropriately sized cannulae.

References

  1. Royal College of Emergency Medicine Emergency Medicine Point of Care Ultrasound curriculum and guidance (2021)
  2. American College of Emergency Physicians Ultrasound Guidelines: Emergency, Point-of-Care and Clinical Ultrasound Guidelines in Medicine (2023)
  3. American College of Surgeons ATLS, 10th edition (2018)
  4. NICE NG39 Major trauma: assessment and initial management (2016)
  5. NICE NG156 Abdominal aortic aneurysm: diagnosis and management (2020)
  6. NICE Technology Appraisal Guidance 49: Ultrasound-guided placement of central venous catheters (2002)

Short checklist (PDF, one page) · Full checklist (PDF)

Educational use. EM Orbit procedure guide — written from published guidance and clinical practice, for education only. Perform procedures only within your training, competence and local policy, under supervision where required. It does not replace your clinical judgement.