Recording a 12-lead ECG
Standardised recording of a resting 12-lead electrocardiogram with correct electrode placement, so that the trace is accurate, comparable and immediately reviewed.
Open in the appFree, works on your phone and offline, with every tool alongside.
Do not forget
- Before In chest pain, record the ECG within 10 minutes of arrival and show it to a decision-making clinician immediately.
- After Do not rely on the machine's automated interpretation to rule out acute myocardial infarction.
Indications
- Chest pain or suspected acute coronary syndrome
- Palpitations, syncope or pre-syncope
- Breathlessness, suspected heart failure or pulmonary embolism
- Suspected electrolyte disturbance (e.g. hyperkalaemia) or drug toxicity (e.g. tricyclic overdose)
- Arrhythmia seen on a monitor, before and after cardioversion, and after return of spontaneous circulation
- Stroke, sepsis, overdose and other presentations where local pathways require a baseline ECG
Contraindications
Absolute
- None — but never delay defibrillation or CPR in cardiac arrest to record a 12-lead ECG
Relative
- Patient refusal
- Burns, wounds or dressings at electrode sites (use the nearest intact skin and annotate the trace)
Equipment
- 12-lead ECG machine, calibrated and with paper
- Ten disposable electrodes (in date, with moist gel)
- Skin preparation: gauze or abrasive tape, alcohol wipe, razor or clippers for chest hair
- Privacy screen or curtain, gown or blanket for modesty
- Chaperone as needed
- Clinical wipes to clean the leads
Key numbers
- Time to ECG in chest pain
- Within 10 minutes of arrival
- Paper speed
- 25 mm/s
- Calibration
- 10 mm/mV
- Diagnostic filter
- 0.05–150 Hz
- V1 / V2
- 4th intercostal space, right / left sternal edge
- V4
- 5th intercostal space, left mid-clavicular line
- V5 / V6
- Same level as V4, anterior / mid-axillary line
Before · Patient preparation
- Identify the patient and enter their details on the machine (name, identifier, date of birth, date and time).
- In chest pain, record the ECG within 10 minutes of arrival and show it to a decision-making clinician immediately.Do not wait until the end of triage.
- Explain the procedure, obtain consent and offer a chaperone.
- Ensure privacy and expose the chest, wrists and ankles; cover the patient where possible.
- Position the patient semi-recumbent at about 45° with arms relaxed at the sides, warm and still.Record the position if different, as posture changes the ECG.
- Ask the patient to relax, breathe normally and not talk during the recording.
Before · Skin preparation
- Clean the electrode sites to remove oil or cream; dry the skin.
- Clip or shave hair at the electrode sites if it prevents good contact.
- Lightly abrade the skin if needed to reduce impedance.
During · Limb electrodes
- RA (red): right arm, on the inner wrist or forearm.
- LA (yellow): left arm, on the inner wrist or forearm.
- LL (green): left leg, above the ankle or on the lower leg.
- RL (black): right leg, above the ankle or on the lower leg.Colours shown are the IEC (European) standard; American AHA colours differ.
- Place limb electrodes distally and symmetrically; if they must be moved to the torso or proximally (e.g. amputation, tremor), annotate the trace.Torso placement alters the limb-lead ECG and is not equivalent.
During · Chest electrodes
- Find the angle of Louis (sternal angle): the second rib attaches here; the space below it is the second intercostal space. Count down to the fourth intercostal space.
- V1: fourth intercostal space at the right sternal edge.
- V2: fourth intercostal space at the left sternal edge.
- V4: fifth intercostal space in the left mid-clavicular line.
- V3: midway between V2 and V4.
- V5: left anterior axillary line at the same horizontal level as V4.
- V6: left mid-axillary line at the same horizontal level as V4 and V5.
- In women, place V4–V6 under the breast rather than on it.Lift the breast gently with the back of the hand, with consent.
During · Recording
- Connect the leads to the correct electrodes; avoid tension and let leads lie without crossing.
- Check settings: paper speed 25 mm/s, calibration 10 mm/mV, and the correct filter settings (diagnostic 0.05–150 Hz where available).
- Check the preview for a clear, stable baseline without artefact; ask the patient to stay still.
- Record the ECG and check the printout for a calibration pulse, correct patient details and all 12 leads.
- Check for lead reversal: a negative P and QRS in lead I with a positive aVR suggests arm lead reversal.
- Record a rhythm strip (long lead II) if an arrhythmia is present.
During · Additional leads
- If inferior ST elevation is present, record right-sided leads (V4R: fifth intercostal space, right mid-clavicular line) to look for right ventricular infarction.
- If posterior infarction is suspected (ST depression in V1–V3), record posterior leads V7–V9 in the fifth intercostal space at the left posterior axillary line, mid-scapular line and paraspinal border.
- Label any non-standard lead positions clearly on the trace.
After · Completion
- Remove the electrodes, help the patient dress and ensure they are comfortable.
- Clean the leads and cables and perform hand hygiene.
After · Review and documentation
- Annotate the ECG with the patient's symptoms at the time (e.g. 'with chest pain'), time recorded and any non-standard positions.
- Hand the ECG to a clinician for review immediately; record who reviewed it and when.
- Do not rely on the machine's automated interpretation to rule out acute myocardial infarction.A clinician must interpret every ECG.
- Compare with previous ECGs and repeat serial ECGs if symptoms persist, change or recur.
- File or upload the ECG to the patient record.
Complications
- Skin irritation from electrodes
- Misdiagnosis from incorrect lead placement or lead reversal
- Delay in recognising ST-elevation myocardial infarction if the ECG is not reviewed promptly
- Loss of dignity if privacy is not maintained
Pitfalls
- Placing V1 and V2 too high (second or third intercostal space), which mimics anterior changes such as poor R-wave progression or Brugada pattern.
- Placing V4–V6 on breast tissue.
- Muscle tremor, movement or electrical interference causing artefact that mimics arrhythmia.
- Recording at the wrong paper speed or calibration.
- Failing to annotate symptoms and positions, so serial ECGs cannot be compared.
In children
Use smaller electrodes; in infants, V4R is often recorded routinely and chest electrodes may need to be placed closer together. Normal values (heart rate, axis, intervals and right ventricular dominance) vary with age; interpret against age-specific ranges.
References
- Society for Cardiological Science and Technology: Clinical guidelines by consensus — Recording a standard 12-lead electrocardiogram (2020)
- NICE CG95 Recent-onset chest pain of suspected cardiac origin (2010, updated 2016)
- ESC Guidelines for the management of acute coronary syndromes (2023)
- Resuscitation Council UK Adult Advanced Life Support Guidelines (2021)
Short checklist (PDF, one page) · Full checklist (PDF)
Educational use. EM Orbit procedure guide — written from published guidance and clinical practice, for education only. Perform procedures only within your training, competence and local policy, under supervision where required. It does not replace your clinical judgement.